Sample id, Measure unit, Alarm – Thermo Fisher Scientific Eutech 2700 Series User Manual

Page 31: Ion setup

Advertising
background image

27

13.

Ion Setup

Use Setup Ion mode to customize this parameter. Press

from the

measurement screen and then press

ENTER when ION is selected.

Sample ID

This is a user selectable number from 1 to 99999.

Incorporating a sample ID to identify one or more

data points is useful to distinguish data that is saved

into memory or sent to a PC or printer. Use the

up/down arrow keys to adjust the values and

left/right arrow keys to move the cursor.

Measure Unit

Select ppm, molar, or mg/L ion concentration units.



Alarm

The 2700 offers a visual and audible alarm to alert

you when the High or Low values that have been

set from this menu are exceeded.

“High Alarm” or “Low Alarm” will blink on the display

if the values are exceeded while simultaneously, a

loud, intermittent beeping sound is heard. The

alarms will continue until the conditions are no

longer met, and will only be active during

measurement mode.