Pressure cal (% saturation mode only), Pressure compensation (% saturation mode only), Measure unit (concentration mode only) – Thermo Fisher Scientific Eutech 2700 Series User Manual

Page 36: 32 pressure cal (% saturation mode only)

Advertising
background image

32

Pressure Cal (% saturation mode only)

The DO 2700 includes a built-in barometer that is

factory calibrated. While frequent pressure

adjustment is not required, adjustments can be

made in the Adjusted Pressure field. The

Measured Pressure is the active pressure value

measured by the DO 2700.
Note: if adjusting, “true” barometric pressure—that

has not been corrected to sea level, must be used.

Pressure reported by weather services are almost always corrected and therefore

are not appropriate for entry in the adjusted pressure field here.

Pressure Compensation (% saturation

mode only)

Select

ENABLE or DISABLE.



Measure Unit (concentration mode only)

Select mg/L or ppm concentration units.


Advertising