Before and after a manual calibration

5) For multi-point calibration repeat with additional standards. Press ESC to

save calibration or press

NEXT to view the calibration report. Calibrate one

point per range, up to 5 points.

Additional manual calibration notes:

A maximum of one calibration point per range can be performed. If multiple

calibration points are used in the same range, the most recent one will replace

the previous one.
When the electrode is replaced, it is best to clear the calibration to the factory

default values.
Rinse or immerse the probe before calibration and between samples with clean

water (deionized water is ideal).
To protect from erroneous calibrations, the allowable tolerance is ±40% of the

factory default value.
For best results always begin with your lowest calibration standard value,

followed by the next lowest, and so on.
Low conductivity standard solutions (less than 20 µS) are unstable and are very

temperature dependent. As a result, reproducible calibration results are

challenging in lowest measurement range #1 (0.00 to 20.0 µS).

Note: If using the 2-cell electrode, the protective probe guard

can be removed temporarily for cleaning however it must be

re-attached during measurement and calibration. Erroneous

results will occur while the probe guard is removed.