Sample id, Offset cal (% saturation only), Pressure unit (% saturation mode only) – Thermo Fisher Scientific Eutech 2700 Series User Manual

Page 35: Dissolved oxygen setup

Advertising
background image

31

15.

Dissolved Oxygen Setup

Use Setup Dissolved Oxygen mode to customize the parameter. Press

from the measurement screen and then press

ENTER when DO% or DO (mg/L)

is selected for percent saturation or concentration modes respectively. Note that

setup parameters differ slightly depending on which is selected.

Sample ID

This is a user selectable number from 1 to 99999.

Incorporating a sample ID to identify one or more

data points is useful to distinguish data that is

saved into memory or sent to a PC or printer. Use

the up/down arrow keys to adjust the values and

left/right arrow keys to move the cursor.

Offset Cal (% saturation only)

Use this feature to offset the meter’s value to

another DO meter. Using the % saturation mode,

observe the reading a sample solution after it has

stabilized. Similarly, observe the reading of the

same sample using another DO meter as a

reference. The probe of the reference meter should

be immersed in the same sample at the same

depth.
The DO 2700 allows ±10% offset adjustment. Note that subsequent user

calibrations will reset the offset adjustment back to the factory default value of 0.0

% offset.

Pressure Unit (% saturation mode only)

Select mmHg or kPa pressure units.

Advertising