Conductivity, TDS, Salinity, & Resistivity Setup

Use Setup mode to customize Conductivity, TDS, Salinity, & Resistivity

parameters.

Press

from the measurement screen and then press

ENTER when

Conductivity, TDS, Salinity, or Resistivity is selected.

Sample ID

This is a user selectable number from 1 to 99999.

Incorporating a sample ID to identify one or more

data points is useful to distinguish data that is

saved into memory or sent to a PC or printer. Use

the up/down arrow keys to adjust the values and

left/right arrow keys to move the cursor.

Cal Method

(conductivity mode only)

The 2700 is capable of automatic or manual

calibration for conductivity, and manual calibration

for TDS, Salinity, and Resistivity. This option

allows you to select

AUTO (automatic) or

MANUAL conductivity calibration.
In the automatic calibration mode, the meter will

automatically select one of (4) conductivity

calibration standard values depending on the range and normalization

temperature being used (see table below).

Range

#

Conductivity

Range

Automatic Calibration Values

Normalization Temperature

25 ºC

20 ºC

r 1

0.00 – 20.00 µS

None

None

r 2

20.1 – 200.0 µS

84 µS

76 µS

r 3

201 – 2000 µS

1413 µS

1278 µS

r 4

2.01 – 20.00 mS

12.88 mS

11.67 mS

r 5

20.1 – 200.0 mS

111.8 mS

102.1 mS

If you will only use one or more of the (4) calibration standards listed here,