17

8.

Temperature Setup

Use Setup Temperature mode to desired temperature units; º C or º F.

Press

from the measurement screen and then press

ENTER when

TEMPERATURE is selected.
Choose º C or º F temperature units of measure.

Automatic Temperature Compensation (ATC) is automatically applied when a

temperature sensor is connected.

Manual Temperature Compensation (MTC) of 25 º C is used if the temperature

sensor is not connected. See

Section 9—Temperature Calibration

to adjust the

MTC temperature value.