Cal points, Pure water coefficient, Temperature coefficient (linear) – Thermo Fisher Scientific Eutech 2700 Series User Manual

Page 24

Advertising
background image

20

automatic calibration is recommended. Otherwise, manual calibration should be

selected.
The factory default is automatic conductivity calibration. Note that automatic

calibration is not available for conductivity range 1.

Cal Points

Use

SINGLE for Single-Point Calibration (SPC) to

apply a single calibration value across all ranges.
Use

MULTI for Multi-Point Calibration (MPC) to

calibrate each range individually. This will restrict

an individual calibration so that it is applied to one

range only. When using multi-point calibration,

perform a calibration in each range that you

expect to use for best results.
The factory default is SPC. SPC or MPC will be shown with the calibration details

in measurement mode.

Pure Water Coefficient

Choose

ENABLE to automatically apply pure

water temperature correction for measurements

below 2

µS (

ultra pure water).

Note: this option does not apply to Salinity mode.

Temperature Coefficient (linear)

The temperature coefficient is the amount of

change in conductivity per degree temperature (%

per ºC). The factory default setting is a

temperature coefficient of 2.1 % per ºC. For most

applications this will provide good results. The

meter allows adjustment from 0.0 to 10.0.
TIP:

Select 0.0% for uncompensated

measurements. The temperature will be measured

by the electrode and displayed in measurement mode—without compensation.