Thermo Fisher Scientific Eutech 2700 Series User Manual

Page 50

Advertising
background image

46

sensor.

DO

Instrument will not

read zero in sodium

sulfite solution

Solution contains oxygen—use fresh

solution.
Replace membrane and recondition

sensor.

Temperature Instrument reads

inaccurate

temperature

Calibrate using water bath or known

accurate thermometer. Faulty

thermistor—repair or replace.

ALL

Main digits are

faded in

measurement

mode &/or are

faded too often.

When “Stability” is enabled, the digits

on the primary reading will be faded

when unstable, becoming solid when

stable. Disable this feature to keep

digits solid at all times or keep this

feature enabled while changing the

“Stability Criteria” to “Slow“

ALL

I forgot my

password

Please send us a written request with

your name, contact information, along

with the instrument serial number; a

temporary password will be provided.

ALL

Memory Full

Often when a timed datalog is used, it

runs until the memory is full. To erase

the stored data, go to System Setup,

CLEAR LOG.

ALL

Reading doesn’t

change, “AR” is

blinking

Auto-Read is active (measurement

locks automatically when stable). Press

MEAS to resume. Select Auto-Read

“DISABLE” if this feature is not desired.

ALL

My problem is not

listed here.

Please contact our authorized dealer or

email us directly with your feedback:

[email protected]