Thermo Fisher Scientific Eutech 2700 Series User Manual

Page 17

Advertising
background image

13

4) The primary display is the un-calibrated measured value. The 2700

automatically searches for and selects the appropriate value from your buffer

group in the secondary display. This value will blink when the ‘Stable’

indicator appears. Selecting

CLR-C will clear the existing calibration.

5) Press ENTER to accept the calibration value

of the measured buffer. When successful the

accepted value will be visible inside of a

beaker icon. Percentage slope is visible after

two or more calibration points have been

completed.

6) Repeat Steps 4 & 5 with additional pH buffers

in any order or press

ESC to escape and save

your calibration. When the specified number of

calibration points is met, the pH calibration

report page is automatically displayed.

To specify a different number of pH calibration

points see

Section 4—Cal Points

.

If the meter is password protected, you will be prompted to

enter a password. See

Section 3—Password

.