Badger Meter B3000 User Manual

Page 16

Advertising
background image

Compact Digital Flow Monitor, Model B3000

Page 16

March 2014

Programming Using Frequency Output Turbine Flow Meters

Each Blancett turbine flow meter is shipped with either a K factor value or frequency data If frequency data is provided, the

data must be converted to a K factor before programming the monitor K factor information, when supplied, can usually be

found on the neck of the flow meter or stamped on the flow meter body The K factor represents the number of pulses per

unit of volume (see

K FACTORS EXPLAINED on page 38

) The K factor is required to program the monitor

Essentials
The B3000 monitor is engineered to provide several levels of programming tailored to the needs of the user The first or

standard level provides access to the most commonly used setup parameters bypassing the more advanced settings The first

level programming is entered by pressing and holding MENU for about one second
The second level or extended settings are accessed by pressing and holding MENU until the extended programming

menu starts
With the standard and solar liquid monitors there is a third level For the most basic unit setup choices the B3000 monitor

employes a simple and advanced setup option accessed through the Rate SU parameter If Simple is selected the rate and total

choices are reduced to the five most common combinations avoiding the need to make unit and interval choices

Liquid Meters

Standard

Solar

Advanced I/O

Basic Functions

Press MENU for about one second and then release

Extended Functions

Press and hold MENU until the extended programming menu starts

Simple Setup

Select Rate SU in the extended functions and choose Simple

Not Applicable

Advanced Setup

Select Rate SU in the extended functions and choose Advanced

Table 1: Display mode selection information

Enter Programming Mode
The programming modes are accessed by pressing MENU for basic functions Extended functions are accessed by pressing

and holding MENU until the first programming parameter appears

2

1

Battery

Indicator

Units

Indicator

Totalizer

Units

Totalizer

Multiplier

Numeric

Values

Function

Figure 15: Programming mode display