Compact Digital Flow Monitor, Model B3000

Page 22

March 2014

Totalizer Pulse Output PulsOut
Basic Function
The PulsOut parameter can be either enabled or disabled When enabled, this output generates a fixed width 30 mS duration,

pulse every time the least significant digit of the totalizer increments The amplitude of the pulse is dependent on the voltage

level of the supply connected to the pulse output and is limited to a maximum 28V DC
The B3000 provides two types of totalizer pulses The basic open drain FET output,

Figure 19

and

Figure 20

, provides a ground

referenced output pulse that swings between about 0 7V DC and V

CC

JP1

JP2

JP3

P1

Freq. In

4-20mA

Iso Total Pluse

TR_B

TR_A

RS485 Gnd

Setpoint 1

Setpoint 2

Gnd

+

+

+

Total Reset

Total Pluse

Signal Gnd

Mag

Pulse

Iso

OC

Low

High

Input

Total P

ulse

Sig
nal

TB1

Open Drain FET

Total Pulse Output

2.2 …10k

Pullup

Resistor

V

CC

Internal

100 mA

Maximum

Freq. In

4-20mA

Iso Total Pluse

+

+

+

Total Reset

OC Total Pluse

Signal Gnd

JP3

Total Pulse

ISO

OC

JP1

Input

Mag

Pulse

TB2

Open Drain FET

Total Pulse Output

2.2…10k

Pullup

Resistor

V

CC

Internal

100 mA

Maximum

Figure 19: Open drain connections (NEMA 4X)

Figure 20: Open drain connections (Ex-Proof)

The isolated pulse output (ISO),

Figure 21

and

Figure 22

, are again an open collector output with the emitter of the transistor

connected to the negative output terminal and is not referenced to ground This output is optically isolated from the input

signal for systems that require a totally isolated output pulse

JP1

JP2

JP3

Input

Total P

ulse

Sig
nal

P1

JP1

JP2

JP3

Input

Total P

ulse

Sig
nal

P1

Freq. In

4-20mA

Iso Total Pluse

TR_B

TR_A

RS485 Gnd

Setpoint 1

Setpoint 2

Gnd

+

+

+

Total Reset

OC Total Pluse

Signal Gnd

TB1

Mag

Pulse

Iso

OC

Low

High

Isolated Output

Total Pulse

Internal

–V

2.2…10k

Pullup

Resistor

V

CC

100 mA

Maximum

Freq. In

4-20mA

Iso Total Pluse

+

+

+

Total Reset

OC Total Pluse

Signal Gnd

JP3

Total Pulse

ISO

OC

JP1

Input

Mag

Pulse

TB2

Internal

Isolated Output

Total Pulse

–V

2.2…10k

Pullup

Resistor

V

CC

100 mA

Maximum

Figure 21: Opto-isolated open collector connections (NEMA 4X)

Figure 22: Opto-isolated open collector connections (Ex-Proof)

Both outputs have a maximum current capacity of 100 mA and require a pullup resistor The value of the pullup resistor is

dependent on the supply voltage and the maximum current required by the load device