Badger Meter B3000 User Manual

Page 47

Advertising
background image

Programming Manual

Page 47

March 2014

Modbus Coil Description

Modbus Coil

Notes

Reset Running Totalizer

1

Forcing this coil ON will reset the running totalizer After reset, the

coil automatically returns to the OFF state

Reset Grand Totalizer

2

Forcing this coil ON will reset both the running totalizer and the

grand totalizer After reset, the coil automatically returns to the OFF

state

3…8

Spares

Alarm Set point 1

9

0 = Set point OFF, 1 = Set point ON

Alarm Set point 2

10

0 = Set point OFF, 1 = Set point ON

11…16

Spares

Table 11: Modbus coil map

Opcode 01 – Read Coil Status
This opcode returns the state of the alarm coils The following Coils are defined:

Coil #

Description

9

Alarm Set point 1

10

Alarm Set point 2

11 and up

Spare

Table 12: Read coil status

Command: <addr><01><00><08><00><02><crc-16>
Reply: <addr><01><01><0x><crc-16>

Opcode 03 – Read Holding Registers
This opcode returns the input holding registers, such as flow rate or totalizer

OTE:

N

Each value must be requested individually Return of a block of registers is not implemented at this time

Example requesting flow rate in floating point format
Command: <addr><03><00><C9><00><02><crc-16>
Reply: <addr><03><02><data><data><crc-16>
Opcode 05 – Force Single Coil
This opcode sets the state of a single coil (digital output) The following Coil Registers are defined:

Coil #

Description

1

Reset Totalizer

2

Grand Totals

3 and up

Spares

Table 13: Force single coil

The transition of coil from 0 to 1 will initiate function This bit is auto reset to 0, so there is no need to set it to 0 after a totalizer

reset command

Command: <addr><05><00><00><FF><00><crc-16>

Reply: <addr><05><00><00><FF><00><crc-16>