Pipe Reducer Requirements

With pipe reducers, a smaller meter can be mounted in larger pipelines . This arrangement may increase low-flow accuracy .
There are no special requirements for standard, concentric, pipe reducers .
Custom fabricated pipe reducers must have an approximate slope angle of 15 degrees to minimize flow disturbances and

excessive loss of head . If this is not possible, install the custom pipe reducers as if they were fittings and install the required

amount of straight pipe

Figure 9: Pipe reducer requirements

Chemical Injection Applications

For water line applications with a chemical injection point, install the meter upstream of the injection point . This eliminates

any meter performance issues .

Figure 10: Chemical injection point upstream of meter

If a meter must be installed downstream of a chemical injection connection, the distance between the meter and the

injection point should be between 50 and 100 feet (15 and 30 meters) . The distance must be long enough to allow the water

or chemical solution to reach the meter in a complete, homogeneous mixture .

Figure 11: Chemical injection point upstream of meter

If the injection point is too close, the meter senses the two different conductivities for each liquid . This will likely result in

inaccurate measurements . The injection method—spaced bursts, continuous stream of drips or liquid or gas—can also affect

downstream readings by the meter .