AMPLIFIER MOUNTING CONFIGURATION OPTIONS

There are two configuration options for mounting the amplifier . There are many options to accommodate a variety of

meter-placement and environmental conditions .

Meter Mount Configuration

The meter mount configuration has the amplifier mounted directly on the detector . This compact, self-contained

configuration minimizes installation wiring .

Remote Mount Configuration

The remote mount configuration places the amplifier at a location away from the fluid flow and detector . This is necessary in

situations where process fluid temperature or the environment exceeds amplifier ratings .
The detector and amplifier are connected by wires, run through conduit, between junction boxes on the detector and the

amplifier . The distance between the detector junction box and amplifier junction box can be up to 150 feet (50 meters) . A

remote mounting bracket is supplied .

Submersible Option

If you are installing the meter in a vault, you should order the remote mount, submersible amplifier option . You must not

install the amplifier inside a vault . This will eliminate any potential problems resulting from humidity or temporary flooding in

the vault .

OTEE:

N

The National Electronics Manufacturer's Association (NEMA) 6P enclosures are constructed for indoor or outdoor use

to provide protection against access to hazardous parts; to provide a degree of protection against ingress of solid

foreign objects and water (hose directed water and the entry of water during prolonged submersion at a limited

depth); that provide an additional level of protection against corrosion and that will be undamaged by the external

formation of ice on the enclosure .

Protection Class

In order to fulfill requirements of the protection class, follow these guidelines:

• BODY SEALS NEED TO BE UNDAMAGED AND IN PROPER CONDITION.
• ALL OF THE BODY SCREWS NEED TO BE FIRMLY SCREWED.
• OUTER DIAMETERS OF THE USED WIRING CABLES MUST CORRESPOND TO CABLE INLETS (FOR M20 Ø 5....13 MM). IN

CASES WHERE CABLE INLET IS NOT USED, PUT ON A DUMMY PLUG.

• TIGHTEN CABLE INLETS.
• IF POSSIBLE, LEAD CABLE AWAY DOWNWARDS. HUMIDITY CANNOT GET INTO CABLE INLET.
• WE NORMALLY DELIVER THE METER IN ACCORDANCE WITH PROTECTION CLASS IP 67. IF YOU HOWEVER REQUIRE A

HIGHER PROTECTION CLASS, THE AMPLIFIER IS TO BE INSTALLED SEPARATELY FROM THE DETECTOR. IF REQUESTED,

WE CAN ALSO DELIVER THE DETECTOR IN IP 68.

Amplifier mounting configuration options

Page 12