Auxiliary power, Remote version – Badger Meter M-Series M1000 User Manual

Page 16

Advertising
background image

Remote Version

CONNECT OR REMOVE THE SIGNAL CONNECTION CABLE ONLY

WHEN THE UNIT HAS BEEN SWITCHED OFF.

Connection to the Measuring Amplifier
1 . Open the cover (see

"Opening the M1000 Cover" on page 14

) .

2 . Push the signal cable through the lower cable inlet .
3 . Connect as shown in

Figure 21

.

4 . Close the cover and tighten the screws and the wire gland .
Connection to the Junction Box
1 . Open the junction box .
2 . Push the signal cable through the upper cable inlet .
3 . Connect as shown in

Figure 21

.

4 . Close the cover and tighten the screws and wire gland .

Figure 21: Remote version connection

Junction Box

M1000

Description

Wire Color

11

C1

Coil 1

Green

12

C2

Coil 2

Yellow

13

CS

Main shield

Yellow/Green

45

E1

Electrode 1

White

44*

ES

Electrode shield

Black

46

E2

Electrode 2

Brown

40

EP

Empty pipe

Pink

44*

ES

Empty pipe shield

Black

* Connections with number 44 are on the same potential .

white (45)

brown (46)

pink (40)

black (44)

black (44)

yellow (12)

green (11)

5 . . . 50 m

green (11)

yellow (12)

pink (40)

white (45)

brown(46)

black (44)

black (44)

yellow/green (13)

yellow/green (13)

Figure 22: Signal cable specification

Power connections

Page 16

July 2014

MAG-UM-00379-EN-02