Meter Setup Menu

Meter Setup

Calibration

Diameter

A

This parameter is set at the factory . In the event the amplifier is replaced,

verify that the pipe diameter matches the installed pipe size .

Detector Factor

A

This parameter is set at the factory . This factor compensates for accuracy

error as a result of the installed detector .
In the event the amplifier is replaced, this parameter must be

reprogrammed with the original detector zero .

Detector Zero

A

This parameter is set at the factory . This parameter compensates for

accuracy error as a result of the installed detector .
If accuracy adjustment of the meter is required, refer to the scale factor .

Amplifier Factor

This parameter is set at the factory and is Read Only . This electronic

calibration factor compensates for accuracy error as a result of the

installed amplifier .

Coil Current

This parameter (coil current to the detector) is set at the factory and is

Read Only . This factor compensates for accuracy error as a result of the

installed amplifier .

Scale Factor

S

Changing the scale factor lets you adjust the meter’s accuracy without disturbing parameters set

by the factory . You can tune the meter to meet changing application requirements in a range of

± 5% (0 .95 to 1 .05) .

Power Line Frequency

S

For optimum operation of the meter, set Power Line Frequency to 50 Hz or 60 Hz in this menu at

operating location .

Excitation Frequency

A

This parameter is set at the factory . This value shows in which frequency the meter’s coils are

operated . Supported frequencies are dependent on the configured power line frequency and

meter’s size .

50 Hz

60 Hz

3 .125 Hz

3 .75 Hz

6 .25 Hz

7 .5 Hz

12 .5 Hz

15 Hz

OTEE:

N

When selecting excitation frequency, make sure to always observe that the ratio in

respect of power frequency is integer .

Empty Pipe Detection

S

ON/OFF

S

When set to ON, an Empty Pipe condition indicates to the outputs

that the meter is not completely filled . When set to OFF, empty pipe

conditions are not detected .

Threshold

S

Threshold value for empty pipe detection .
For liquids with lower conductivity or long cables the threshold value

must be increased . The actual value can be monitored in the next menu

“measured” .

Measured

Measures the real empty pipe value . This parameter is Read Only .

M1000 main menu programming options

Page 21

July 2014

MAG-UM-00379-EN-02