Meter setup menu – Badger Meter M-Series M1000 User Manual

Page 22

Advertising
background image

Measurement Menu

Measurement

Flow Unit

U

Flow Units let you select among the Flow Units listed below . Flow units are automatically converted

into the selected unit .

Display

Flow Unit

Display

Flow Unit

L/s

Liters/Second

gal/s

Gallons/Second

L/min

Liters/Minute

gal/min

Gallons/Minute

L/h

Liters/Hour

gal/h

Gallons/Hour

m

3

/s

Cubic Meters/Second

MG/d

MillionGallons/Day

m

3

/min

Cubic Meters/Minute

IG/s

ImperialGallons/Second

m

3

/h

Cubic Meters/Hour

IG/min

ImperialGallons/Minute

ft

3

/s

Cubic Feet/Second

IG/h

ImperialGallons/Hour

ft

3

/m

Cubic Feet/Minute

oz/min

Ounce/Minute

ft

3

/h

Cubic Feet/Hour

bbl/min

Barrel/Minute

Totalizer Unit

U

This parameter establishes the units of measure for the totalizers .

Display

Totalizer Unit

Display

Totalizer Unit

L

Liters

MG

Million Gallons

hL

Hectoliter

IG

Imperial Gallons

m

3

Cubic Meters

bbl

Barrel

ft

3

Cubic Feet

oz

Fluid Ounces

gal

U .S . Gallons

ac/ft

Acre per foot

Full Scale Flow

S

This parameter sets the maximum flow the system is expected to measure . This parameter has

influence on other system parameters like analog output or low flow cutoff .
In terms of flow velocity, the meter’s range is 0 .3…12 m/sec .
The full scale flow is valid for both flow directions .

OTEE:

N

Note: If the flow rate exceeds the full scale setting, an error message indicates that the

configured full scale range has been exceeded .

Low Flow Cutoff

U

Low Flow Cutoff defines the threshold at which flow measurement will be forced to zero . The cutoff

value can be from 0% to 10% of the full scale flow . Increasing this threshold will help prevent false

readings during “no flow” conditions possibly caused by vibrations or liquid fluctuations .

Flow Direction

S

Flow direction lets you set the meter to measure forward flow only (unidirectional) or both forward and

reverse flow (bidirectional) .
Unidirectional means that the flow is totalized in only one direction . The flow direction is indicated by

the arrow printed on the detector label . In this mode, the two totalizers T1+ and T2+ can be used as

totalizers and resettable day counter .
Bidirectional means the flow is totalized in both directions . The totalizer T1+ and T2+ registers forward

flow and the totalizer T1– and T2– in reverse flow direction . The net totalizer T1N and T2N shows the

difference between T+ and T– .
A change of the flow direction can be signalized by the digital outputs .

M1000 main menu programming options