Inputs/Outputs

Digital Outputs

(continued on next

page)

Pulse Width

S

This parameter establishes the “On” duration of the transmitted pulse . The

configurable range is 0…2000 ms . If 0 ms is configured, pulse width is

automatically adapted depending on pulse frequency (pulse/pause ratio 1:1) .

During the configuration the program checks if pulses/unit and pulse width are

in accordance with full scale defined, if not an error alarm is displayed . In case of

an error alarm, scale, pulse width or full scale need to be adapted .

Pulse/Unit

S

The Pulses/Unit parameter lets you set how many pulses per unit of measure

will be transmitted . The maximum output frequency of 10,000 pulses/sec .

(10 kHZ) must not be exceeded .

For example, assuming the unit of measure is gallons:

• Setting the Pulses/Unit to 1 will transmit 1 pulse every gallon

• Setting the Pulses/Unit to 0 .01 will transmit 1 pulse every 100 gallons

You must configure pulses/unit if the function of the selected output is to be

forward, reverse or AMR pulse .

Frequency

S

This parameter establishes to define the digital output as frequency output . The

full scale frequency configurable range is 0 .01…10,000 Hz .

Set Min./Max.

S

The Flow Set Point (min, max) establishes as a percentage of full scale flow,

the threshold at which the output alarm will be activated . You can freely select

thresholds in 1% steps . Flow rates below/above the threshold will activate the

output alarm .

Preset Amount

S

Preset amount lets you set the reset value for the associated PS totalizer when

the digital input is set to Batch Reset . You can configure preset amount in the

adjusted volume unit .

Preset amount is counted down from the configured value to 0 and a digital

output shows that the preset amount has been reached .

Out 1 Function

Out 2 Function

S

(continued on next

page)

From the sub-menus Out 1 Function and Out 2 Function, you can configure

functional operation of the 2 digital outputs . For example, you can select

“forward pulse” for the digital output and define the pulses-per-totalizer unit via

“pulse scale” .

The two outputs can be operated as open collector passively or actively .

Passive output:

Active output (if analog output is not used):

M1000 main menu programming options