Measurement menu – Badger Meter M-Series M1000 User Manual

Page 23

Advertising
background image

Measurement

Filter

S

Median

S

The Median Filter (MDN) reduces noise on the measuring signal . The filter

level can be adjusted from 7 up to 13 or switched off .

Moving Average

S

Moving average filter (MAV) smooths out short-term fluctuations . The value

can be adjusted from 1 to 200 measuring periods .
The delay is calculated:

Delay [s] = ( MAV – 1) x T
The time (T) is given by the adjusted excitation frequency of the meter .

Excitation frequency [Hz]

T = Time for filter delay (s)

15

0 .03333

12 .5

0 .040

7 .5

0 .06666

6 .25

0 .080

3 .75

0 .13333

3 .125

0 .160

For example: MAV = 20 and the excitation frequency is 6 .25 Hz

means T=0 .08 s, the delay is 1 .52 s .

Display

S

Moving average filter smooths out short-term fluctuations only for the

display . The value can be adjusted from 1 to 200 measuring periods .
For calculation of the delay see Moving Average above .

M1000 main menu programming options

Page 23

July 2014

MAG-UM-00379-EN-02