Badger Meter M-Series M1000 User Manual

Page 26

Advertising
background image

Inputs/Outputs

Digital Outputs
(continued)

Out 1 Function

Out 2 Function

S

(continued)

Solid-State Relay

The solid-state relay is functionally linked with Output 2 . See the table below .

The following functions can be selected for the Outputs 1 to 2 as well as for

the solid-state relay . The solid-state relay function is linked with the function of

Output 2 .

Function

Meaning

Out1

Out2 /

Solid-

State

Relay

Off

Digital output is switched off .

X

X

Test

Used only for the Verification Device .

X

X

Comparator

TBD

X

X

Empty Pipe

Error

Indicates when a pipe is empty .

X

X

Error Alarm

Indicates a meter error condition .

X

X

Forward

Pulses

Generates pulses during forward flow

conditions .

X

X

Reverse Pulses

Generates pulses during reverse flow

conditions .

X

X

Direction

Indicates current flow direction .

X

X

Loopback

Shows the status of the digital input .

X

X

Min ./Max .

Alarm

Establishes, as a percentage of full scale flow,

the threshold at which the output alarm will

be activated . Flow rates below or above the

threshold will activate the output alarm .

X

X

Frequency

TBD

X

X

Rotary

Encoder

TBD

X

X

Preset

Indicates when a preset batch amount has

been realized .

X

X

Out Type 1

S

This parameter lets you set the output switch to normally open or normally

closed . If normally open is selected, the output switch is open (no current) when

the output is inactive, and closed (current flows) when the output is active .

If normally closed is selected, the output switch is closed (current flows) when

the output is inactive, and open (no current) when the output is active .

Out Type 2

S

This parameter lets you set the output switch to normally open or normally

closed . If normally open is selected, the output switch is open (no current) when

the output is inactive, and closed (current flows) when the output is active .

If normally closed is selected, the output switch is closed (current flows) when

the output is inactive, and open (no current) when the output is active .

Simulation

S

Flow Simulation provides analog and digital output simulation based on a percentage of the full

scale flow in cases where no real flow is occurring . The range of simulation is –100…100% in steps

of 10% of the full scale flow . This function still remains active once you have left the menu . It is

necessary to set it to OFF to deactivate it . If the simulation is still active, the letter “S” will be displayed

in the measuring mode .

Totals Menu

Totals

Clear T2

U

The unidirectional Totalizer T2 is reset within the menu manager .

M1000 main menu programming options

Page 26