Badger Meter M-Series M1000 User Manual

Page 24

Advertising
background image

Input/Outputs Menu

Input/Outputs

Analog Output

Range

S

This parameter establishes the range of the analog output signal: 0…100% (= full

scale) . The following current ranges are available to you:

Current Output

0…20 mA
4…20 mA
0…10 mA

Analog output active

Analog output passive

OTEE:

N

When an error message is displayed, the current is set according the

programing of the Alarm Mode below .

When you select bidirectional operation, you can signal flow direction via

digital outputs .

Alarm Mode

S

This parameter configures the behavior of the analog output during alarm

conditions . Three options exist for this parameter: OFF, LOW and HIGH .
OFF: Analog signal is based on flow rate and always within the configured range .
LOW: During alarm conditions, the analog signal will be 2 mA less than the

configured lower range . (only on 4…20 mA range) .
HIGH: During alarm conditions, the analog signal will be 2 mA more than the

configured upper range .
For example, if the analog range is 4…20 mA and the alarm mode is set to HIGH,

then during a full scale flow alarm condition, the analog output current will be

22 mA .

Digital Input

S

Digital input lets you reset totalizers (Remote reset), or interrupt flow measurement (PosZeroReturn) .

• Remote Reset – Clears totalizer T2 (unidirectional)

• Pos Zero Return – Forces flow rate to zero (does not totalize)

Input switching is provided by applying an external potential of 5…30V DC

or by an internal voltage source of 24V DC (Analog output if not used) .

M1000 main menu programming options

Page 24

July 2014

MAG-UM-00379-EN-02