Miscellaneous

Log

Off, On and Preset

Power Up

The number of times that the unit has been powered on .

Settling Time

Measures settling of coils and must be less than one quarter of the excitation period . Zero ms if no

detector is connected .

Language

This parameter allows changing the current language . English is the default setting .
The following languages are supported:

German (Deutsch), Czech (Cestina), Spanish (Espanol), French (Francais), Russian (России),

Italian (Italiano) .

Date

Set date of the system in the format day, month, year [DD .MM .YY] used for data logging .

Time

Set time of the system in the format hour, minutes, seconds [HH .MM .SS] used for data logging .

EEPROM

Delete all data logging information from the EEPROM .

Polar Voltage

Measure electrode polarizing voltage in ± V (only for service purpose) .

Display Rotation The Display can be rotated to 0°, 90°, 180° and 270° .
Contrast

The contrast of the display can be adjusted between 14 (low) and 49 (high) .

Datalog Period

The data logging period can be adjusted to the following increments:
• Every 15 min / 1 h / 6 h / 12 h / 24 h
There is a 500 kB memory with about 30,000 data records for data logging available . The logging

capacity is as following (Unidirectional mode):

Time Period

Duration

15 min

up to 312 days

1 h

up to 1250 days

6 h

up to 20 years

12 h

up to 40 years

24 h

up to 80 years

Startup, configuration, and error events that are logged can reduce the data logging capacity . Logging

in bidirectional mode reduces the logging capacity by about 40% .
The logging information can be downloaded by a PC program, which can be ordered with the

following part number: 67354-010 .

Information Menu

Info

Serial Number

Serial number of the electronic board .

Version

Software version of the device .

Compilation Date Date of the software version .
OPT CRC

Checksum of the software update .

APP CRC

Checksum of the application .