PIN Menu

PIN

Control

A

Activate and deactivate the PIN .

User

U

Users logged in with this PIN have access to all user levels . Users at this level do not have access to

Service or Admin functions .

Service

S

Users logged in with this PIN have access to both service and user-level procedures . Users at this level

do not have access to Admin functions .

Admin

A

Users logged in with this PIN have access to all procedures . Users at this level have full access to the

meter .

Login Screen

Login

Login

This screen is available only after the password protection (PIN Control) has been set to ON . Enter your

6-digit password .

MAINTENANCE

Mandatory, routine or scheduled maintenance should not be required for the M1000 Mag Meter electronics or flow tube after

proper installation . However, some occurrences may require personnel to perform the following:

Flow tube and electrode cleaning

Circuit board replacement

DO NOT CLEAN COMPONENTS INSIDE THE AMPLIFIER OR JUNCTION BOX.

Cleaning the Flow Tube and Electrode

At times flow tube, electrodes, amplifier/junction box housings and the amplifier window may need periodic cleaning,

depending on process fluid properties, fluid flow rate and surrounding environment .
Clean the flow tube and electrodes by following the material handling and cleaning procedures documented in Material

Safety Data Sheet (MSDS) guidelines for the products(s) that were in contact with the flow tube and electrodes .
Should flow tube and/or electrode cleaning become necessary:
1 . Disconnect detector from pipeline .
2 . Clean electrodes according to MSDS guidelines .
3 . Reconnect detector to pipeline .

Maintenance

Page 29

July 2014

MAG-UM-00379-EN-02