KM9106 Operators Manual

26

Parameter

Controls Options

REF NOx

Select using or key and

NOx = SUM

NOx = NO

NOx = NO2

There are three ways of displaying the value of NOx when the values are converted to
mg/m3. Local authorities may require a certain calculation. The options are as
follows:

NOx = SUM calculates the mg/m3 figure individually from the NO and NO2
values and then adds them together.

NOx = NO adds the ppm values together and then converts to NO equivalent.

NOx = NO2 adds the ppm values together and then converts to NO2
equivalent.

On power up the unit will default to NOx = SUM