UEi Test Instruments KM9106/P User Manual

Page 35

Advertising
background image

KM9106 Operators Manual

35

The above example will send the Time and Oxygen reading to the printout every two
minutes.

To stop the Quintox Auto printing or logging, select Auto print OFF above and return
to the main displays.

7.2.1 Standard Printout - Parameter Options Used:

LINE

NUMBER

PARAMETER

USED:

(MAX LINES: 40) & STATUS:

1

- - - - - - - - - - - -

PRINTOUT START

2

BLANK LINE

3

***KANE-MAY***

MANUFACTURER ID

4

**KM QUINTOX**

ANALYSER ID

5

BLANK LINE

6

DATE 02-01-97

DATE

7

TIME 18 :14:35

TIME

8

BLANK LINE

9

NATURAL GAS

FUEL SELECTED

10

BLANK LINE

11

02 % ......20.9

FLUE GAS 0

2

CONTENT

12

CO ppm .....0.0

FLUE GAS CO CONTENT

13

Prs mbar ....0.05

PRESSURE MEASUREMENT

14

EFF %.....FAULT

COMBUSTION EFF CALC

15

BLANK

LINE

16

XAIR O2>20%

FLUE GAS EXCESS AIR

17

C02 %.......0.0

FLUE GAS CO

2

CALC (CO2m% FLUE GAS

CO

2

MEASURE)

18

CO/CO2 R....0.00

CO/CO

2

RATIO

19

PI % .....0.00

POISON INDEX

20

NO ppm.....0

FLUE GAS NO CONTENT

21

NO2 .....0

FLUE GAS NO

2

CONTENT

22

NOx ppm.....0

NOX CALCULATION

23

SO2 ppm.....0.0

FLUE GAS S0

2

CONTENT

24

CxHy % ...... 0.0

HYDROCARBON READING

25

BLANK LINE

26

NETT C 0.0

NET FLUE GAS TEMP

27

FLUE C 21.0

ACTUAL FLUE GAS TEMP

28

INLT NOT

FITTED

BOILER INLET TEMP

29

AMBIENT C 16.9

AMBIENT AIR TEMP

30

- - - - - - - - - - - -

DOTTED LINE

Not all parameters are used. See the master list on the next page if more are required.