6.6 Oxygen Calibration

If used over long periods the Oxygen sensor may drift slightly and for maximum
accuracy may require resetting.

TO RESET OXYGEN SENSOR

With the flue probe sampling fresh air and the flue temperature reading less than
50

°C/122°F, or the temperature plug disconnected from the instrument.

Press the

-- -- -- -- -- -- -- -- -- -- -- -- --

SET

-- -- -- -- -- -- -- -- -- -- -- -- --

Pressing the key displays


Change to YES

SET O2 = 20.9%

NO

Press to set 20.9% Oxygen.

NATURAL GAS

PRESS ENTER KEY

6.7 Toxic sensor zero

The CO and other optional toxic sensors can be reset to zero if they drift. This may
happen if the sensor is taken to very high concentrations for long periods of time or
over-ranged.

Press twice

-- -- -- -- -- -- -- -- -- -- -- -- --

SET

-- -- -- -- -- -- -- -- -- -- -- -- --

ENTER SERVICE CODE

-- -- -- -- > 0 0 0 0 < -- -- -- --

Press four times

CALIBRATE SENSOR

NO