KM9106 Operators Manual

27

6.5 Inlet Temperature

The Quintox uses as default the internal ambient sensor when calculating the Net
temperature. If an optional inlet probe is fitted then INLET is used in the calculation.

As an alternative to both of the above, the inlet air entering the boiler can be measured
using the flue probe.

NOTE:

The probe must not be inserted into the flue until the INLET temperature
has been set. If resetting the inlet temperature after performing a
combustion test ensure the tip of the probe is at the air temperature


TO SET INLET TEMPERATURE

With the flue probe connected to the FLUE temperature connector.

Press the

-- -- -- -- -- -- -- -- -- -- -- -- --

SET

-- -- -- -- -- -- -- -- -- -- -- -- --

Pressing the key displays


Change to YES

SET INLET TEMP

NO

Position the tip of the flue probe near the
Air inlet of the boiler and when the reading
is stable press

PLACE PROBE BY

AIR INLET

-- -- -- -- -- -- -- -- -- -- -- -- --

FLUE . . . 30.0


The temperature measured by the flue probe will be set in the Quintox.

Press

to check the reading.


INLET

SET/CAL

ENTER