CO/CO2 R :

The CO/CO2 ratio, is the ratio of measured CO divided by CO2.


It gives an indication of the following :-


How good a gas sample the instrument is reading.

How clean the boiler is running.


For example : A new or clean domestic boiler will display a ratio
of less than 0.004, a unit in need of cleaning 0.004-0.008 and a
unit in need of major overhaul will show greater than 0.008.


This only shows a reading when a combustion test is being
carried out. 0.0000 is displayed while in fresh air.


P INDEX :

The CO/CO2 ratio expressed as a percentage %, called the

‘Poison Index”

i.e. P INDEX % = 100 x CO/CO2. 0.00 is displayed while in

fresh air.


XAIR % :

Excess air calculated from the measured oxygen and type of fuel
used.

Displays reading during a combustion test. O2 > 20% is
displayed while in fresh air.


Prs :

Flue draught pressure reading. Displayed when pressure sensor
fitted. See section 4.5 for taking a reading and Section 6.4 for
changing for scale.


NO:

Nitric Oxide reading in ppm or mg/m3. Displayed when Nitric
Oxide sensor fitted. Reading can also be referenced to oxygen
ppmn or mg/m3n.


NO2:

Nitrogen Dioxide reading in ppm or mg/m3. Displayed when
Nitrogen Dioxide sensor fitted. Reading can also be referenced to
oxygen ppmn or mg/m3n.


NOx :

Calculated total Nitric oxides displayed in ppm or mg/m3.

For more details on NOx calculation see Section 6.4 Scales.
Reading can also be referenced to oxygen ppmn or mg/m3n.


SO2 :

Sulphur Dioxide reading in ppm or mg/m3. Displayed when
Sulphur Dioxide sensor fitted. Reading can also be referenced to
oxygen ppmn or mg/m3n.