Select yes, Set zero ? no, Select sensor co – UEi Test Instruments KM9106/P User Manual

Page 29: Set zero level co instability 5 7 calibrating 77

Advertising
background image

KM9106 Operators Manual

29


CALIBRATE
SENSOR

Select using or key and

YES

NO

Select YES

SET ZERO ?

NO

Parameter

Controls Options

SET ZERO

Select using or key and

YES

NO

Select YES

SELECT SENSOR

CO -------- . . -17

Parameter

Controls Options

SELECT
SENSOR

Select using or key and

LIST OF

FITTED

SENSORS

DISPLAYED


In this example we have selected the CO sensor. The pump will now run if it has been
turned off, this is to draw fresh air through the instrument to allow the sensors to be
zeroed.

Ignore the reading next to the sensor


Ensure the instability reads zero

SET ZERO LEVEL

CO -------- . . -17

INSTABILITY 5
7

CALIBRATING 77


Once the instability has reached zero press to accept the zero calibration
and return to the main display.

If the instability does not reach zero then the instrument will show FAULT. Contact
Kane International or Authorised Distributor for advice.