Konica Minolta bizhub PRESS C1060 User Manual

Page 59

Advertising
background image

c. Press the measuring button.

After approximately 1 second the device beeps.

d. While pressing the measuring button, slide the instrument

along the ruler from left to right.

Note: Try to slide the instrument at a constant speed. For an optimal
slide speed, the instrument chaetal take approximately 2 to 4
seconds with the protection glass attached (5 to 7 seconds with the
polarization filter attached) to move from one end of the ruler to the
other. if the speed is too fast or too slow, an error may occur.

e. When the specimen measuring port comes to the position

where it touches the other end of the ruler's opening on the
right, release the measuring button.
When you release the measuring button the device beeps.

If the scan is successful, a check mark

appears in the

wizard above the icon of the color you have just scanned,
and an arrow appears below the icon of the next color to
scan. The instructions in the wizard change accordingly.

If the scan is not successful a red and white icon
appears above the color you have just scanned indicating
that the scan was not successful and that you should re-
scan that row,.

When all of the separation rows have been successfully
scanned, a check mark appears above each of the color
separation icons.

Creating calibration tables with the Konica Minolta FD-5 spectrodensitometer

49

Advertising
This manual is related to the following products: