Location – HMC Electronics MDL Series BK Precision DC Electronic Load System Module User Manual

Page 25

Advertising
background image

25

The following figures show examples of how channels are assigned when single-channel and dual-
channel modules are installed.

Figure 5 - Channel Number Order When Installing Four Single-Channel Modules

Figure 6 – Channel Number Order When Installing Two Single-Channel Modules and Two Dual-Channel Modules


Note : Mainframe extension module channels are numbered 11-18.


Location

The operating temperature of the MDL Series DC Electronic Load is 0 to 40 ℃. A fan cools the
electronic load by drawing air through both the top and front, and then exhausting it out the back.
Therefore, the electronic load must be installed in a location that allows sufficient space on the
front and back of the unit for adequate air circulation. Minimum clearances for bench operation
are 2 inches from the top and 3 inches from the front and back. If there are radiator fans in your