HMC Electronics MDL Series BK Precision DC Electronic Load System Module User Manual

Page 53

Advertising
background image

53

A/B Transient Operation Tutorial

Panel Operations

VFD Display

1.

Power on the electronic load.

Self-test

2.

Press  keys to select the channel to be edited. In this example, we

select channel 1.

CH01 CC OFF
Vdc = 0.0000V
Adc = 0.0000A
Wdc = 0.00W

3.

Press Setup to enter the channel setup menu. Select <Mode> and

press Enter to change the operating mode to CC mode. Press Enter to
confirm.

CH01
Mode=CC
RANGE=HIGH
Iset=9.000A

4.

Press  key to select <Range> setting and press Enter. Then press 

key to select LOW RANGE and press Enter to confirm.

CH01
HIGH RANGE
LOW RANGE

5.

Press  key to select the rise slope setting <∫> and press Enter. For

1A/us, input 1 and press Enter to confirm.

CH01
Rise Speed Set
= 1.00A/Us

6.

Press  key to select the fall slope setting <∫> and press Enter. For

2A/us, input 2 and press Enter to confirm.

CH01
Fall Speed Set
= 2.00A/Us

7.

Press  key to select A level setting <TRANa> and press Enter. For

10A, input 10 and press Enter to confirm.

CH01
Transition A
Level = 10.00A

8.

Press  key to select A width setting <Ta> and press Enter. For 0.002s,

input .002 and press Enter to confirm.

CH01
Tran. A Width =
0.002S

9.

Press  key to select B level setting <TRANb> and press Enter. For 2A,

input 2 and press Enter to confirm.

CH01
Transition B
Level = 2.00A

10.

Press  key to select B width setting <Tb> and press Enter. For 0.003s,

input .003 and press Enter to confirm.

CH01
Tran. B Width =
0.003S

11.

Press  key to select transient operation mode <Tmode> and press

Enter. Select Continuous and press Enter key to confirm.

CH01
Continuous
Pulse
Toggle

Documentation Provided By HMC Electronics

33 Springdale Ave. Canton, MA 02021

http://www.hmcelectronics.com

(800) 482-4440