Remote Sense Function

When working in CC, CV, and CR mode, if the electronic load consumes a very large current, it will
cause a voltage drop in the leads between the connected device and terminals of the electronic
load. In order to ensure testing accuracy, the electronic load provides a pair of remote sensing
terminals in the rear panel where users can sense the output terminal voltage of the connected
device. Users should set the electronic load in REMOTE SENSE mode before using this function. On
the rear terminals, SENSE (+) and SENSE (–) are the remote sensing inputs. By eliminating the effect
of the voltage drop in the load leads, remote sensing provides greater accuracy by allowing the
electronic load to regulate directly at the source's output terminals (see Figure 12).

External Analog Control

You can control the current setting of the electronic load in CC mode using the external analog
programming terminal, pin 7 and pin 8. A 0-10 V input signal will simulate 0 – full scale of the
electronic load to regulate the input current of the electronic load (10V indicates the full range of
electronic load’s current rating).

Automatic Test

The automatic test function of the MDL Series electronic load is useful for simulating various tests
and allows the user to edit up to 10 program files. Each file has 10 steps and up to 100 steps can be
edited and saved into the EEPROM. Convenient for production environments, automatic test can
cascade sequences across multiple channels and allows setting of P/F criteria.

Setting Up Test Files

This is a step-by-step tutorial on how to set up a test file.

Panel Operations

VFD Display

1.

Power on the electronic load.

Self-test

2.

Select the channel you want to edit and then edit every group’s

step. For this example, we will select channel 3 and channel 5
below. Press  keys to switch channel to 3.

CH03 CC OFF
Vdc=0.0000V
Adc=0.0000A
Wdc=0.00W