HMC Electronics MDL Series BK Precision DC Electronic Load System Module User Manual

Page 56

Advertising
background image

56

2.

Press  keys to select the channel to be edited.

CH01 CC ON
Vdc=0.0000V
Adc=0.0000A
Wdc=0.00W

3.

Press Shift + ⑦ to enter System menu. Press  key to select

<Trigger source> and press Enter. Select <Manual> and press Enter
to confirm.

MANUAL
EXTERNAL
HOLD
BUS
TIMER

4.

If electronic load is on, press On/Off key to turn off the electronic

load’s input.

CH01 CC OFF
Vdc=0.0000V
Adc=0.0000A
Wdc=0.00W

5.

Press Shift + ⑧ to enter Configuration menu. Press  key to select

<List> and press Enter. Press  key to select <Edit List> and press
Enter.

FUNCTION MODE
RECALL LIST
EDIT LIST
EXIT

6.

Press  key to select high range or low range and press Enter to

confirm.

High Range
Low Range

7.

Input number of steps in List. Input value and press Enter to confirm.

List file step
=

5

8.

Set first step’s current level. Input value and press Enter key to

confirm.

Step 001 level
=

1A

9.

Set first step’s slew rate. Input value and press Enter key to confirm.

Step 001 rate
=

0.5A/us

10.

Set first step’s width time. Input value and press Enter key to

confirm.

Step 001 width
=

6S

11.

Set second step’s current level. Input value and press Enter key to

confirm.

Step 002 level
=

2A

12.

Set second step’s slew rate. Input value and press Enter key to

confirm.

Step 002 rate
= 0.1A/us