List Mode Programming Example
If in remote control mode, you can refer to the following example to edit List operation (refer to
MDL Series Programming Guide for more information).

Command

Function

LIST:RANGe 40

Sets List range

LIST:COUNT 10000

Sets List cycle

LIST:STEP 2

Sets List steps

LIST:LEVEL 1,5

Sets List step 1 level

LIST:SLEW 1,2

Sets List step 1 slew rate

LIST:WIDTH 1,1

Sets List step 1 width

LIST:LEVEL 2,1

Sets List step 2 level

LIST:SLEW 2,2

Sets List step 2 slew rate

LIST:WIDTH 2,2

Sets List step 2 width

LIST:SAV 2

Saves List to file 2

LIST:RCL 2

Recalls List file 2

FUNCTION:MODE LIST

Sets List mode function

TRIG:SOUR BUS

Sets trigger source

*TRG

Runs List file


Note : This example programs a list on a MDL200 module. For other models, make sure all
range, level, slew, and transient width parameters are within the modules’ specified limits.