Examples, Error messages, 5 examples 2.6 error messages – INFICON LDS3000 Modular Leak Detector Interface description User Manual

Page 15

Advertising
background image

ASCII Protocol

15

2.5

Examples

2.6

Error Messages

:CALSniff

calibration factor sniff

521

R/S

:CALVac

calibration factor vacuum

520

R/S

*SERVICE

:READBuffer

Read service buffer

1300 ..
1310

R

*STARTFLA
SH

Flash-Update starten

2619

S

*RST

:FACTORY

Sets all parameters to factory default

1161

S

:CALHistory

Clears calibration history

1161

S

:ERRORHistory

Clears error history

1161

S

Re

la

te

s to

L

D

cmd. no.

Read /

Set

Command

answer

*stat? (CR)

MEAS (CR)

mode

*status? (CR)

MEAS (CR)

mode

*read? (CR)

2.876E-7 (CR)

leak rate according to
programmed unit

*read:pa*m3/s? (CR)

2.876E-6 (CR)

leak rate in a different unit

*start (CR)

OK (CR)

start measurement

*conf:trig1? (CR)

1.0E-9 (CR)

retrieve trigger 1

*conf:trig1 2.0E-9 (CR)

OK (CR)

set trigger 1

Message

Meaning

OK

command completed

E01

wrong command start (no „*“)

E02

illegal blank

E03

command word 1 illegal

E04

command word 2 illegal

E05

command word 3 illegal

E06

control by RS232 not enabled

E07

argument faulty

E08

no data available

E09

error buffer overflow

E10

command invalid

E11

query not allowed

E12

only query allowed

E13

not yet implemented