Examples, Error messages, 5 examples 2.6 error messages – INFICON LDS3000 Modular Leak Detector Interface description User Manual
Page 15

ASCII Protocol
15
2.5
Examples
2.6
Error Messages
:CALSniff
calibration factor sniff
521
R/S
:CALVac
calibration factor vacuum
520
R/S
*SERVICE
:READBuffer
Read service buffer
1300 ..
1310
R
*STARTFLA
SH
Flash-Update starten
2619
S
*RST
:FACTORY
Sets all parameters to factory default
1161
S
:CALHistory
Clears calibration history
1161
S
:ERRORHistory
Clears error history
1161
S
Re
la
te
s to
L
D
cmd. no.
Read /
Set
Command
answer
*stat? (CR)
MEAS (CR)
mode
*status? (CR)
MEAS (CR)
mode
*read? (CR)
2.876E-7 (CR)
leak rate according to
programmed unit
*read:pa*m3/s? (CR)
2.876E-6 (CR)
leak rate in a different unit
*start (CR)
OK (CR)
start measurement
*conf:trig1? (CR)
1.0E-9 (CR)
retrieve trigger 1
*conf:trig1 2.0E-9 (CR)
OK (CR)
set trigger 1
Message
Meaning
OK
command completed
E01
wrong command start (no „*“)
E02
illegal blank
E03
command word 1 illegal
E04
command word 2 illegal
E05
command word 3 illegal
E06
control by RS232 not enabled
E07
argument faulty
E08
no data available
E09
error buffer overflow
E10
command invalid
E11
query not allowed
E12
only query allowed
E13
not yet implemented