Hardware configuration for profibus 6.5.1, Assignment of the profibus address, 5 hardware configuration for profibus – INFICON LDS3000 Modular Leak Detector Interface description User Manual
Page 52: 1 assignment of the profibus address

52
Fieldbus Communication
Fieldbus supports all commands from LD protocol, except the commands in the
following list:
6.5
Hardware Configuration for Profibus
Sequence of the data words (slots) must be:
• Output at first, inputs at second.
• One or two words are accessible at once.
• Output and inputs must have the same memory start address.
6.5.1
Assignment of the PROFIBUS Address
The PROFIBUS address can be assigned via CU1000 or via the hardware configu-
ration tool of the PLC.
To assign the PROFIBUS address via CU1000 select
► "Main Menu -> Settings -> Set up -> Interfaces -> Bus module -> Address".
To assign the PROFIBUS address via hardware configuration tool of the PLC
► refer to the documentation af your PLC.
If you use a Siemens Step 7 you can also
26
Interface protocol (read only)
27
Used interface
275
Cal history
287
Error history
288
TMP error history
294
Text of error number
406
Serial number leak detector
407
Serial number MSB
450
Date + Time [YMDhms]
1161
Parameter reset
1300
Service buffer ion current
1301
Service buffer pressure 1
1302
Service buffer emis current
1303
Service buffer anode voltage
1304
Service buffer cathode voltage
1305
Service buffer heater power
1306
Service buffer leakrate
1307
Service buffer TMP mode
1308
Service buffer TMP speed
1309
Service buffer emission mode
1310
Service buffer sensor 3
2619
Start flash update
2594
Compatibility mode
2661
Set maintenance
2662
Maintenance done