Fieldbus Communication

Fieldbus supports all commands from LD protocol, except the commands in the
following list:

6.5

Hardware Configuration for Profibus

Sequence of the data words (slots) must be:

• Output at first, inputs at second.

• One or two words are accessible at once.

• Output and inputs must have the same memory start address.

6.5.1

Assignment of the PROFIBUS Address

The PROFIBUS address can be assigned via CU1000 or via the hardware configu-
ration tool of the PLC.

To assign the PROFIBUS address via CU1000 select

► "Main Menu -> Settings -> Set up -> Interfaces -> Bus module -> Address".

To assign the PROFIBUS address via hardware configuration tool of the PLC

► refer to the documentation af your PLC.

If you use a Siemens Step 7 you can also

26

Interface protocol (read only)

27

Used interface

275

Cal history

287

Error history

288

TMP error history

294

Text of error number

406

Serial number leak detector

407

Serial number MSB

450

Date + Time [YMDhms]

1161

Parameter reset

1300

Service buffer ion current

1301

Service buffer pressure 1

1302

Service buffer emis current

1303

Service buffer anode voltage

1304

Service buffer cathode voltage

1305

Service buffer heater power

1306

Service buffer leakrate

1307

Service buffer TMP mode

1308

Service buffer TMP speed

1309

Service buffer emission mode

1310

Service buffer sensor 3

2619

Start flash update

2594

Compatibility mode

2661

Set maintenance

2662

Maintenance done