LDS1000 Protocol

Q/U14

Correction factor for SNIFF

e.g.: 1.0E0

Q/U16

Operating mode for ext. CAL The
setting is not saved when switching
off the mains power.

0 - with autotune
1- dyn. CAL without autotune

Q/U19

Request for CAL (Enable CAL
message for a temperature
difference of 5° C).

0 - off
1 - on

Q/U20

Mass of the gas which is detected
in the mass spectrometer

2 , 3, 4
e.g.: 4

Q/U21 Date

e.g.: 24.Nov04
Abbreviations for the months:
Jan May Sep Feb Jun Oct Mar
Jul Nov Apr Aug Dec

Q/U22 Time

e.g.:

14:40:07

Q/U24

Unit (unit of measurement for
pressure and leak rate in VAC
and SNIFF)

ppm and g/a is not available for
VAC

0 - mbar and mbar l/s
1 - Pa and Pa m

3

/s

2 - atm and atm cc/s
3 - mbar and g/a
4 - mbar and ppm
5 - Torr and Tor l/s

Q/U27

Leak rate of the internal
calibrated leak (always in mbar
l/s

e.g.: 1.0E-7
9.9E-1 for not available

Q/U28

leak rate of the external
calibrated leak

e.g.: 1.0E-5
9.9E-1 for not available

Q/U31

Number of suppressed decades 0 - 1 to 2 decades

1 - 2 to 3 decades
2 - 3 to 4 decades
3 - 2 decades
4 – complete value
5 - 19/20 of value

Q/U32

Zero suppression when START

0 - off
1 - on

U45

Compatibility
Mode

2 – LDS2010-Mode
3 – LDS3000-Mode

Q/U56 Factor

500G

- 15G

Q/U57

MSV anode potential for masse 2
in volts

e.g.: 890

Command

Meaning

Representation