6 fieldbus communication, Preface, Setup – INFICON LDS3000 Modular Leak Detector Interface description User Manual

Page 49: Write process data (plc-> leak detector), 6fieldbus communication, 1 preface, 2 setup, 3 process data mapping for cyclic data transfer, 1 write process data (plc-> leak detector)

Advertising
background image

Fieldbus Communication

49

6

Fieldbus Communication

6.1

Preface

In order to use Fieldbus Communication with LDS3000, you need an INFICON Bus-
Module BM1000 connected to the I/O port of the LDS3000.

When setting up the PROFIBUS communication you need to use the GSD file
provided by INFICON.

The following file contains the communication characteristics for the PROFIBUS:
HMSB1811.GSD

You can download the file at
http://www.anybus.de/

6.2

Setup

► Select the „Bus modul“ at the control unit (CU1000): M

ENU

> S

ETTINGS

> S

ETUP

>

I

NTERFACES

> A

CCESSORY

> D

EVICE

SEL

. > B

US

> OK.

► Select the field bus at the control unit (CU1000): M

ENU

> S

ETTINGS

> S

ETUP

>

I

NTERFACES

> B

US

M

ODULE

>

ADDRESS

.

Information Attention: This value do not come into effekt until a restart of the leak

detector (power off / power on)!

6.3

Process Data Mapping for Cyclic Data Transfer

6.3.1

Write Process Data (PLC-> Leak Detector)

This data is sent periodically from the programmable logic controller to the leak
detector:

Byte Bit Name

Meaning

Similar to
PLC Input

Similar to
RS232 ASCII
cmd.

Similar to
RS232 LD
cmd.

1

0

(not used)

1

Zero

Transition 0 -> 1: 0x02 = Zero on
Transition 1 -> 0: 0x00 = Zero off

ZERO

*ZERO

6

2

Clear

Transition 0 -> 1: 0x04=Clears errors and
warnings

Clear

*CLS

5

3

Start/Stop

Transition 0 -> 1: 0x08= Start
Transition 1 -> 0: 0x00= Stop

Start / Stop

*START /
*STOP

1, 2

4

CAL intern

Transition to 0: 0x00 = Cancel internal
calibration
Transition to 1: 0x10 = Start internal
calibration

CAL intern

*CAL:INT

4

5

6

CAL extern

Transition to 0: 0x00 = Cancel external or dyn.
calibration
Transition to 1: 0x40 = Start external or. Dyn.
calibration
Transition to 2: 0x80 = Acknowledge closed
test leak

CAL extern /
CAL dynamic

*CAL:EXT

4

7