Binary Interface Protocol

Example:

SET Trigger 2 to 1.2E-7mbarl/s

HOST

 LDS2010:

LDS2010

 HOST

GET Trigger 2 in mbarl/s

HOST

 LDS2010:

82

83

GetZeroMode

SetZeroMode

Choice zero function

0=2-3 Decades;
1=1-2 Decades;
2=19/20 of valuet;

3=2 Decades;
4=3-4Decades
5=complete value

84

85

GetFacSniff

SetFacSniff

Read sniff factor

[float]

92

93

GetUnit

SetUnit

unit read / set

Byte 0: LR-vac; Byte 1:
LR-sniff; Byte 2: pressure

0-mbar / mbarl/s; 1-Pa /
Pam³/s; 2-atm / atmcc/s;
3-Torr / Torrl/s

only for LR-sniff: 4-ppm;
5-g/a

99

GetLr

Leak rate

Unit (0-mbar*l/s,
1-Pa*m³/s, .2-atmcc/s,
3-Torrl/s;

In sniff mode
additionally 4-ppm,
5-g/a)

[float]

Nr

Name

Description

Parameter

Data

5

10

57

2

0

52

0

217

89

176

0x05

0x0A 0x39

0x02

0x00 0x34 0x00 0xD9 0x59 0xB0

Start

Len

Cmd

Para0

Para1 Data Data Data Data Checksum

Trigger Trig. 2 mbarl/s 1.2E-7

(4-Byte float)

3

57

60

0x03

0x39

0x3C

Len

Cmd

Checksum