LDS1000 Protocol

19

S14

ZERO status

"Zero"
0 - no correction
1 - a constant leak rate is suppressed

S18

CAL request status

See command Q/U 19
0 - no request
1 - request is present (temperature difference of
5°)

Serviceinformationen, die bei Rückfragen oder im Fehlerfall zur Lösung eines
Problems beitragen können
S30

software version

e.g.:1.00

S31

Serial number

xxxxxxxxxxxxxxx

S32

Operating hours counter xxxxxx

S35

Valve position

xy (always 2 characters)

"1" valve open

"0" valve closed

Byte x Valve for calibrated leak

Byte y Sniffer valve

S39

Status of the remote
contol inputs

See TH ??? Chapter ???.

xxxxxxx (always 7 signs) (Byte 0 first)

Byte 0: Input 7

Byte 1: Input 6

Byte 2: Input 5

Byte 3: Input 4

Byte 4: Input 3

Byte 5: Input 2

Byte 6: Input 1

Byte 7: always 0

S41

Preamplifier

Amplification of the preamplifier can be changed
through F26 … F30. xy

x: Status: 0 - auto, 1 - manuell

y: Amplification: 0 - 13M; 1 - 470M; 2- 15G; 3 -
0,5T

S42

Turbo pump

xxxxx (Byte 0 first)

Byte 0: speed too low

Byte 1: speed too high

Byte 2: always 0

Byte 3: FAIL converter ("1"-Error)

Byte 4: running up/ acceleration