INFICON LDS3000 Modular Leak Detector Interface description User Manual

Page 38

Advertising
background image

38

LD Protocol

402

192

Leak rate filter R/W UINT8

0, 1 ,2

0 = 2-zone
1 = I•CAL
2 = Fixed

403

193

Leak rate
threshold for
averaging time
[mbar*l/s]

R/W FLOAT

1E-11, 1E-10,
9.9E3

Leak rate threshold for averaging time in
mbar*l/s. Below this value the averaging
time is 10,24s. Above this value the
averaging time is 160ms.

406

196

Serial number
leak detector

R

CHAR[11]

Serial number of the complete leak
detector

407

197

Serial number
MSB

R

CHAR[11]

Serial number of the MSB

408

198

Serial number
IO modul

R

CHAR[11]

Serial number of the IO modul

409

199

Zero with start R/W UINT8

0, 0 ,1

Zero with Start
0 = OFF, 1 = ON

410

19A

Zero mode

R/W UINT8

0, 0, 5

unterdrückte Dekaden:
0 = suppress all
1 = 1 -2 decades background
suppression
2 = 2 -3 decades background
suppression
3 = 2 decades background suppression
4 = 3-4 decades background
suppression
5 = 19/20 of the raw signal background
suppression

411

19B

Zero time

R/W UINT16

0 , 5 , 30

Update interval for offset value if leakrate
signal is negative.
Resolution 0,1 s (50 = 5,0 s)

412

19C

Zero Sniffer
Key Enable

R/W UINT8

0, 1, 1

0 = zero key disabled
1 = zero key enabled

419

1A3

Calibration
request enable

R/W UINT8

0, 0, 1

0 = Calibration request disabled
1 = Calibration request enabled

430

1AE

Pressure unit

R/W UINT8

0, 0, 3

Pressure unit
mbar = 0
Pa = 1
atm = 2
Torr = 3

431

1AF

Leak rate unit
vacuum

R/W UINT8

0, 0, 2

Leak rate unit vacuum
0 - mbarl/s
1 - Pam

3

/s

2 - Atm ccs
3 - Torrl/s
4 - ppm
5 - g/a

Comm
and dez

Com
mand
hex

Name

R/W

Data type

Min-, Def.-, Max-
value

Meaning

Advertising