Binary Interface Protocol
27
LDS2010
HOST
4.3
Error messages
232
RS232Invalid Temporary not allowed (example CAL during run up)
240
RS232Cmd Command existiert nicht
243
RS232Len Numbrer or length of parameters faulty
244
RS232Para Parameter out of acceptable range
252
RS232Start First character wrong (unlike 0x05)
253
RS232Checksum Transmited and calculated Checksumme unlike
254
RS232Timeout Timeout (Transmission of a command not completed after
500 msec )
255
RS232Buffer Bufferoverflow (Overflow of the Receive-Buffers)
5
6
56
2
0
69
0x05
0x06
0x38
0x02
0x00
0x45
Start
Len
Cmd
Para0
Para1
Checksum
Trigger
Trig. 2
mbarl/s
7
57
52
0
217
89
166
0x07
0x39
0x34
0x00
0xD9
0x59
0xA6
Len
Cmd
Data
Data
Data
Data
Checksum
1.2E-7 (4-Byte float)