INFICON LDS3000 Modular Leak Detector Interface description User Manual

Page 29

Advertising
background image

LD Protocol

29

Cmd H/L: Command: Command-specifier

Command name in plain text

• 7-Bit ASCII, only printable characters (0x20 and 0x7E)

• Always in English

• Units in square brackets

Command info

Bit 15 ... 13 Meaning

High Nibble (Hex) Comments

000

Read value

0

001

Write value

2

010

Read lower limit value

4

Min values also
defined for read
commands.

011

Read upper limit value

6

Max values also
defined for read
commands.

100

Read default value

8

Def values also
defined for read
commands.

101

Read command name in
plain text

A

Please refer to
chapter “Command
name in plain text”
below.

110

Read command info

C

Please refer to table
“Command info”
below

111

not used

E

1. Byte

Data type (see table “Data types”)

2. Byte

Number of array elements:
0 = no data, no array
1 = data, no array
2 ... 255 = array

3. Byte

Bit 0: 1 = Reading allowed, 0 = Reading not allowed
Bit 1: 1 = Writing allowed, 0 = Writing not allowed
Bit 2 ... 7: always 0 (not used)