Entry of instrument settings – INFICON LDS3000 Modular Leak Detector Interface description User Manual

Page 21

Advertising
background image

LDS1000 Protocol

21

3.2.4

Entry of Instrument Settings

The settings of parameters in the control modus "RS232" may be changed via the
command "U" for update when the jumper XJ1 has been set to RS232. The parame-
ters may be output via the serial interface through the command "Q" for query.
Forexample, "U0, 1.0E-4<cr>" changes the level for the first trigger to 1.0x 10-4.

Through "Q0<cr>" the trigger level can be read.

The settings are each explained in the Technical Handbook jina50e1-a.

In order to use the commands U51 to U66 the password needs to be entered.

G8

Electronics temperature (ELTA) in
°C

e.g.: 23.5

G9

Amplifier temperature (EVSTA) in
°C

e.g.: 29,2

G10

Anode potential (MIAP) in volts.

e.g.: 457

G11

Cathode potential (MIKP) in volts. e.g.: 378

G12

Suppressor potential (MISP) in
volts.

e.g.: 330

G13

Anode-Cathode potential (MIAKP)
in volts.

e.g.: 79

G19

Speed of the turbopump (TMP) in
Hz.

e.g.: 1048

Command

Meaning

Representation

Command

Meaning

Representation

Q/U0

Trigger 1 in current unit

e.g.: 1.0E-5

Q/U1

Trigger 2 in current unit

e.g.: 1.0E-5

Q/U2

Trigger 3 in current unit

e.g.: 1.0E-5

Q/U3

Trigger 4 in current unit

e.g.: 1.0E-5

Q4

Output the operating mode

x, y (always 2 signs)
X: 0 – SPS, 1 - RS232
Y: 0 – VAC. 1 - SNIFF

U4

Select operating mode This setting
is not saved when switching the
mains power off.

0 - VAC
1 - SNIFF

Q/U7

Sensitivity Threshold. Leak

rate in

current unit at which the sensitivity
(averaging time) is

switched over.

e.g.: 1.0E-10

Q/U8

Zero time in seconds (period of
time for which the leak rate signal
must remain below the saved
background level until the saved
background level itself is
corrected).

e.g.: 5

Q10

Always 0

Q11

Limit-Low in current unit

e.g.: 1.0E-8

Q12

Limit-HIGH in current unit

e.g.: 1.0E4

Q/U13

Machine factor for VAC

e.g.: 1.0E0