Main functions, 1 main functions – INFICON LDS3000 Modular Leak Detector Interface description User Manual

Page 17

Advertising
background image

LDS1000 Protocol

17

Through the reset character <ESC> (27d or 1Bh) without <cr> the interface of the
LDS 1000 may be reset back to a defined state. A received string which might be
processed at that moment is erased and its processing is terminated. Receiving of
the <ESC> character is acknowledged by "OK<cr>" (In the case of the "TERMINAL"
program from Microsoft the character "O" is not displayed when the local echo is on).
Thereafter, the interface is ready to receive. Through this character its is easily
possible to check whether or not the data link has been properly installed.

3.2.1

Main functions

X1*)

Calibration: In the STANDBY mode, the internal calibration is started.

In the MEASURE mode, the external calibration is started. The status of the external
calibration may be queried through S12. Sequence of commands for external calib-
ration:

The internal calibration process is run automatically. There after, the LDS3000 will
be in the STANDBY mode.

Command

Meaning

Reply
from the
LDS1000

Equivalent to key
or PLC input

LR

Leak rate, date, time, output status

START

Start measurement mode,
suppress the background which
was measured upon operating
START

OK

MEAS active

STOP

Stop the measurement mode,
display the current background
level

OK

MEAS inactive

ZERO

ZERO mode on, suppress the
background which was measured
upon operating ZERO

OK

ZERO active

ZERO OFF

ZERO mode off, display the
background which was measured
upon operating ZERO

OK

ZERO inactive

CAL

x1)

Internal/external calibration

OK

CAL

CLEAR

Interrupt calibration/erase error
status

OK

CLEAR active

Command

Reply
from the
LDS1000

Meaning

1

START

OK

The LDS1000 enters the measurement
mode, the calibrated leak must be opened,
wait until the signal has stabilised.

2

CAL

OK

External calibration is being started.

3

S12

1

External calibration is running.

4

S12

2

Calibrated leak must be closed, wait until
the signal is stable.

5

CAL

OK

Calibration is continued.

6

S12

0

Calibration complete, the LDS1000 is in
the measurement mode, the instrument is
running in the MEASUREMENT mode.