Request for measurement data, 3 request for measurement data – INFICON LDS3000 Modular Leak Detector Interface description User Manual

Page 20

Advertising
background image

20

LDS1000 Protocol

3.2.3

Request for Measurement Data

Measurement data can be queried through the command G for "GET".

S43

Emission control

xxxxx (Byte 0 first)

Byte 0: Status number

Byte 1: Nominal status

0- off, 1 - Standby, 2 -on

Byte 2: Actual status

0 - off, 1 - Standby, 2 -on

Byte 3: Cathode

1 - Cathode 1, 2 - Cathode 2

S51

Calibration factor M4
Vacuum

e.g.: 7.492E-13

S52

Calibration M4 Sniff

e.g.: 7.492E-13

S70

Output the number of the
current interface error

"ok", if no error is present.

S72

Output the number of the
current error message
(except interface errors)

e.g.: ER53 12.Oct. 11:50

S73

Output the number of the
wrong parameter

"ok", if no error is present.

Meaning

Representation

Command

Meaning

Representation

G6

Forevacuum pressure (PV) in
volts (1000 mbar: 10.0V).

e.g.: 02.629

G7

Preamplifier signal (EVS) in
volts.

e.g.: 01.456

G8

Electronics temperature (ELTA)
in °C

e.g.: 23.5

G9

Amplifier temperature (EVSTA)
in °C

e.g.: 29,2

G10

Anode potential (MIAP) in volts. e.g.: 457

G11

Cathode potential (MIKP) in
volts.

e.g.: 378

G12

Suppressor potential (MISP) in
volts.

e.g.: 330

G13

Anode-Cathode potential
(MIAKP) in volts.

e.g.: 79

G19

Speed of the turbopump (TMP)
in Hz.

e.g.: 1048

Measurement data for servicing:

G6

Forevacuum pressure (PV) in volts
(1000 mbar: 10.0V).

e.g.: 02.629

G7

Preamplifier signal (EVS) in volts. e.g.: 01.456