Error messages, 5 error messages – INFICON LDS3000 Modular Leak Detector Interface description User Manual

Page 48

Advertising
background image

48

LD Protocol

5.5

Error messages

Telegram error handling

• Slave discards all characters until it receives a STX as telegram start identifier.

• Slave does not generate an error message, if address is not correct.

• Slave reports CRC errors with error message 1 (CRC failure)

• Slave reports length errors with error message 2 (Illegal telegram length) or 11

(Data length is not correct for the command)

To prevent the response from colliding with the next request, the slaves do not
respond in case of a timeout.

Error numbers (for Stw: Bit 15 to 1)

In case of error: STX, LEN, Stw, Cmd and one Data-Byte (with error number) sent

1

CRC-failure

2

Illegal telegram lenght

10

command doesn't exist

11

Data length is not correct for the command

12

Read not allowed

13

Write not allowed

14

Array-Index out of range or missing

20

Control actually not allowed with this interface

21

Password not OK

22

Command actually not allowed (e.g. calibration during Run-Up)

30

Data not in range

31

No data available

Advertising