INFICON LDS3000 Modular Leak Detector Interface description User Manual

Page 32

Advertising
background image

32

LD Protocol

10

A

TMP nominal
status

R/W UINT8

TMP nominal status
0 = off
1 = on

11

B

Calibration
acknowledge

W

UINT8

1 = Continue calibration
0 = cancel calibration

12

C

Open/close int.
testleak

R/W UINT8

0 = close 1 = open
incl. Emission monitoring (less sensitive)
internal calibration will overwrite the state

128

80

Leak rate [sel.
unit]

R

FLOAT

Leak rate in selected unit

129

81

Leak rate
[mbar*l/s]

R

FLOAT

Leak rate in mbar*l/s

130

82

Internal
pressure 1 [sel.
unit]

R

FLOAT

Pressure p1 in selected unit

131

83

Internal
pressure 1
[mbar]

R

FLOAT

Pressure p1 in mbar

132

84

Internal
Pressure 2 [sel.
unit]

R

FLOAT

Pressure p2 in selected unit

133

85

Internal
Pressure 2
[mbar]

R

FLOAT

Pressure p2

134

86

Pressure
sensor 3

R

FLOAT

Sensor (0...10 V). Config via commands
2630,2634,2638

135

87

Pressure
sensor 4

R

FLOAT

Sensor (0...20 mA) config via commands
2632,2636,2639

138

8A

TMP actual
rotation speed
[Hz]

R

UINT16

TMP actual rotation speed

139

8B

TMP power [W] R

FLOAT

TMP power in Watt as reportet by TMP
controller

140

8C

TMP operation
hours [h]

R

UINT32

TMP operation hours

141

8D

Frequency
converter
operation hours
[h]

R

UINT32

Frequency converter operation hours [h]

142

8E

Leak detector
operation hours

R

UINT32

Leak detector operation hours

143

8F

TMP
temperature
bottom [deg. C]

R

FLOAT

TMP temperatur bottom [deg. C]

144

90

TMP
temperature
electronic [deg.
C]

R

FLOAT

TMP temperatur electronic [deg. C]

Comm
and dez

Com
mand
hex

Name

R/W

Data type

Min-, Def.-, Max-
value

Meaning