INFICON LDS3000 Modular Leak Detector Interface description User Manual
Page 32

32
LD Protocol
10
A
TMP nominal
status
R/W UINT8
TMP nominal status
0 = off
1 = on
11
B
Calibration
acknowledge
W
UINT8
1 = Continue calibration
0 = cancel calibration
12
C
Open/close int.
testleak
R/W UINT8
0 = close 1 = open
incl. Emission monitoring (less sensitive)
internal calibration will overwrite the state
128
80
Leak rate [sel.
unit]
R
FLOAT
Leak rate in selected unit
129
81
Leak rate
[mbar*l/s]
R
FLOAT
Leak rate in mbar*l/s
130
82
Internal
pressure 1 [sel.
unit]
R
FLOAT
Pressure p1 in selected unit
131
83
Internal
pressure 1
[mbar]
R
FLOAT
Pressure p1 in mbar
132
84
Internal
Pressure 2 [sel.
unit]
R
FLOAT
Pressure p2 in selected unit
133
85
Internal
Pressure 2
[mbar]
R
FLOAT
Pressure p2
134
86
Pressure
sensor 3
R
FLOAT
Sensor (0...10 V). Config via commands
2630,2634,2638
135
87
Pressure
sensor 4
R
FLOAT
Sensor (0...20 mA) config via commands
2632,2636,2639
138
8A
TMP actual
rotation speed
[Hz]
R
UINT16
TMP actual rotation speed
139
8B
TMP power [W] R
FLOAT
TMP power in Watt as reportet by TMP
controller
140
8C
TMP operation
hours [h]
R
UINT32
TMP operation hours
141
8D
Frequency
converter
operation hours
[h]
R
UINT32
Frequency converter operation hours [h]
142
8E
Leak detector
operation hours
R
UINT32
Leak detector operation hours
143
8F
TMP
temperature
bottom [deg. C]
R
FLOAT
TMP temperatur bottom [deg. C]
144
90
TMP
temperature
electronic [deg.
C]
R
FLOAT
TMP temperatur electronic [deg. C]
Comm
and dez
Com
mand
hex
Name
R/W
Data type
Min-, Def.-, Max-
value
Meaning