5. Send monitor status (Code #4)

Instructs the monitor to send the status data. The description of the status data is
given in

Table 10-6

.

Example: To instruct the monitor to send the status data the computer would send:

Chr$(255)+Chr$(254)+Chr$(1)+Chr$(4)+Chr$(0)+Chr$(251)

Table 10-6 Status Data

Parameter Name

Length (bytes)

Range

Active Film #

1

(1-100)

Active Sensor #

1

(1-2)

Sensor #1 Health

1

(0-99 or 255=Failed)

Sensor #2 Health

1

(0-99 or 255=Failed)

Source & Sensor Shutter

1

(Bit6 = Source, Bit7 =
Sensor.
0=Closed,1=Open)

Display #

1

(0=Rate&Thk, 1=Freq,
2=Test)

I/O Address

1

(0-31)

Crystal Type

1

(0=2.5MHz, 1=3.0MHz,
2=5.0MHz, 3=6.0MHz,
4=9.0MHz, 5=10.0MHz)

Simulate Mode

1

(Bit 4, 0=Off, 1=On)

Failure Register

1

Bit0 = Ram Fail Bit1 =
EPROM Fail Bit2 = Film
Data Fail Bit3 = Sensor
Input Fail Bit4 = Not Used
Bit5 = Power Fail Bit6 =
Sensor Card Fail

Total 10 bytes